Pyromol-E-Test

Detection of protein residue in endoscope channels

Pyromol-Test is based on a dye binding method which is also used in clinical chemistry. Contamination in an endoscope contains protein which can be detected at a low level. For routine monitoring and validation of cleaning efficacy according to EN ISO 15883.

Function

  • Easy to use test kit
  • Quick detection of insoluble protein residue
  • Colour change to blue indicates a contaminated channel
  • Dye binding method for a detection limit of 1μg
  • No interference with glutaraldehyde or peracetic acid

Evaluation

Pyromol-E will detect protein residue after manual or automated reprocessing. A blue spot on the swab indicates protein residue in the tested channel. A large amount of protein can discolor the entire test.

Contaminated instruments are a hygiene risk and need to be rejected. Instruments need to be free of any residue!

Article No. Ø in mm Length Pack (pcs.)
3FRIN555141 1,0 2 m 6
3FRIN555142 1,7 2 m 6
3FRIN555143 2,8 2 m 6
3FRIN555144 3,8 2 m 6
3FRIN555145 5,0 2 m 6